Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8000A: (1 year) CAL VER /5500 DATE: 01-Aug-96 AUTHOR: Fluke Corporation REVISION: $Revision: 1.3 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 41 NUMBER OF LINES: 132 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 8000A instruction manual (PN 347906). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 90 day specifications of the 5500A are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F W 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP Connect the UUT to an ac power source. 1.005 DISP Turn the UUT on. 1.005 DISP Warm-up time: 1 hour. 1.005 DISP Ambient temperature: 22C - 25C. 1.005 DISP Relative humidity: less than 70%. 1.006 DISP Connect the 5500A and the UUT as follows. 1.006 DISP [32] 5500A NORMAL HI to UUT V-Ohm 1.006 DISP [32] 5500A NORMAL LO to UUT COMMON 1.007 HEAD {ZERO OFFSET} 1.008 RNG 200mV 1.009 5500 0.00V S 2W 1.010 EVAL Does the UUT read 00.0, flashing +/-00.1? 2.001 EVAL Does the UUT read <= +/-01.0? 3.001 HEAD {DC VOLTAGE} 3.002 HEAD DC VOLTAGE: {200mV Range} 3.003 5500 200 190.0mV 0.3U 2W 4.001 5500 200 -190.0mV 0.3U 2W 5.001 HEAD DC VOLTAGE: {2V Range} 5.002 5500 2 1.900V 0.003U 2W 6.001 5500 2 -1.900V 0.003U 2W 7.001 HEAD DC VOLTAGE: {20V Range} 7.002 5500 20 19.00V 0.03U 2W 8.001 HEAD DC VOLTAGE: {200V Range} 8.002 5500 200 190.0V 0.3U 2W 9.001 HEAD DC VOLTAGE: {1200V Range} 9.002 5500 1200 1000V 2U 2W 10.001 HEAD {AC VOLTAGE} 10.002 HEAD AC VOLTAGE: {1200V Range} 10.003 5500 1200 1000V 7U 100H SI 2W 11.001 HEAD AC VOLTAGE: {200V Range} 11.002 5500 200 190.0V 2.1U 20kH SI 2W 12.001 5500 200 190.0V 2.1U 10kH SI 2W 13.001 HEAD AC VOLTAGE: {20V Range} 13.002 5500 20 19.00V 0.12U 10kH SI 2W 14.001 5500 20 19.00V 0.21U 20kH SI 2W 15.001 HEAD AC VOLTAGE: {2V Range} 15.002 5500 2 1.900V 0.012U 100H SI 2W 16.001 5500 2 1.900V 0.021U 20kH SI 2W 17.001 HEAD AC VOLTAGE: {200mV Range} 17.002 5500 200 190.0mV 2.1U 20kH SI 2W 18.001 5500 200 190.0mV 1.2U 100H SI 2W 19.001 HEAD {RESISTANCE} 19.002 HEAD RESISTANCE: {20 MOhm Range} 19.003 5500 20 19.00MZ 0.11U 2W 20.001 HEAD RESISTANCE: {2000 kOhm Range} 20.002 5500 2000 1900kZ 5U 2W 21.001 HEAD RESISTANCE: {200 kOhm Range} 21.002 5500 200 190.0kZ 0.5U 2W 22.001 HEAD RESISTANCE: {20 kOhm Range} 22.002 DISP Connect the 5500A and the UUT as follows. 22.002 DISP [32] 5500A NORMAL HI to UUT V-Ohm 22.002 DISP [32] 5500A NORMAL LO to UUT COMMON 22.002 DISP [32] 5500A AUX HI to UUT V-Ohm 22.002 DISP [32] 5500A AUX LO to UUT COMMON 22.003 5500 20 19.00kZ 0.05U CW 23.001 HEAD RESISTANCE: {2 kOhm Range} 23.002 5500 2 1.900kZ 0.005U CW 24.001 HEAD RESISTANCE: {200 Ohm Range} 24.002 5500 200 190.0Z 0.5U CW 25.001 HEAD {DC CURRENT} 25.002 DISP Connect the 5500A and the UUT as follows. 25.002 DISP [32] 5500A AUX HI to UUT mA 25.002 DISP [32] 5500A AUX LO to UUT COMMON 25.003 HEAD DC CURRENT: {200uA Range} 25.004 5500 200 190.0uA 0.7U 2W 26.001 5500 200 -190.0uA 0.7U 2W 27.001 HEAD DC CURRENT: {2mA Range} 27.002 5500 2 1.900mA 0.007U 2W 28.001 HEAD DC CURRENT: {20mA Range} 28.002 5500 20 19.00mA 0.07U 2W 29.001 HEAD DC CURRENT: {200mA Range} 29.002 5500 200 190.0mA 0.7U 2W 30.001 HEAD DC CURRENT: {2000mA Range} 30.002 5500 2000 1900mA 7U 2W 31.001 HEAD {AC CURRENT} 31.002 HEAD AC CURRENT: {2000mA Range} 31.003 5500 2000 1900mA 21U 100H SI 2W 32.001 5500 2000 1900mA 21U 3kH SI 2W 33.001 HEAD AC CURRENT: {200mA Range} 33.002 5500 200 190.0mA 2.1U 100H SI 2W 34.001 5500 200 190.0mA 2.1U 10kH SI 2W 35.001 HEAD AC CURRENT: {20mA Range} 35.002 5500 20 19.00mA 0.21U 100H SI 2W 36.001 5500 20 19.00mA 0.21U 10kH SI 2W 37.001 HEAD AC CURRENT: {2mA Range} 37.002 5500 2 1.900mA 0.021U 100H SI 2W 38.001 5500 2 1.900mA 0.021U 10kH SI 2W 39.001 HEAD AC CURRENT: {200uA Range} 39.002 5500 200 190.0uA 2.1U 100H SI 2W 40.001 5500 200 190.0uA 2.1U 10kH SI 2W 41.001 END